UK-Swindon: Microscopios electrónicos de barrido
Scanning electron microscopes. The Authority has a requirement for a scanning electron microscope which has the following core essential criteria:
(1) Non diffusion-pumped vacuum system (turbo-molecular preferred) backed by a rotary pump;
(2) High Vacuum and Variable pressure operation. Variable pressure range from 1 to 400 Pa.;
(3) Chamber size to allow for specimens of around 100mm in height and 200mm in diameter;
(4) Image resolution 4nm or better @ 30kV;
(5) Image resolution 20nm or better @ 1kV;
(6) Low end magnification X10 or less;
(7) High end magnification X500,000 or more;
(8) Secondary electrons (SE) specimen topography detection incorporated;
(9) Backscattered electrons (BSE) chemical composition contrast detection incorporated;
(10) Motorised 5 axis stage;
(11) Dual channel operation;
(12) Liquid Nitrogen (LN)-free cryogenic cooling;
(13) Delivery, installation, commissioning and on site training;
(14) 5-year through life support (first year covered by manufacturers warranty); a minimum of one site maintenance visit per year.
Hardware:
1. Tungsten emission system, variable pressure air admittance to 400 Pa, 5-axis motorised stage with software joystick;
2. Two 17" or better High contrast, colour flat panels;
3. Dual Joystick control device for manual control of motorised stage;
4. Lens Mounted five segment advanced silicon diode backscatter detector, sensitive to 1keV;
5. Dual channel operation which allows the user to display 2 live electron detector images at 1 024 x 768 pixel resolution on to 2 separate screens simultaneously;
6. Chamberscope with infrared (IR) illumination displayed to full screen;
7. Centre feature & centre point stage.
The following are highly desirable options:
1. Energy-dispersive X-ray spectroscopy (EDX) elemental analysis detection incorporated, energy resolution of 133eV or better with Silicon drift detector (SDD) non-cryogenic detector with:
(a) Point analysis;
(b) Line scan analysis;
(c) Area scan analysis;
(d) Area mapping of distributions and inclusions.
Qualitative and quantitative output (e) EDX Microanalysis System.
2. Electron backscatter diffraction (EBSD);
3. Wavelength dispersive X-ray spectroscopy (WDX) Companies must also confirm that they hold, or are working toward third party accreditation to ISO 9001:2008 and ISO 15632:2002 and provide a copy of the same, which must also include the scope of accreditation and must also include coverage of the capability which is the subject of this advert. A pre qualification questionnaire may be required though the Authority reserves the right to proceed to the invitation to tender stage without conducting a PQQ. Companies that wish to express an interest in this requirement are required to provide their full company name, postal address, contact name, telephone number and e-mail address.
Dirección postal: 511 Coldhams Lane
Población/ciudad: Cambridge
Código postal: CB1 3JS
País: GBR
El contenido de esta página constituye exclusivamente un servicio adicional, y no tiene efectos jurídicos. Las instituciones de la UE no asumen responsabilidad alguna por su contenido. Las versiones oficiales de los anuncios de licitación pertinentes son aquellas publicadas en el suplemento del Diario Oficial de la Unión Europea y disponibles en TED. Se puede acceder a esos textos oficiales directamente mediante los enlaces integrados en esta página. Para más información, véase la declaración relativa a la explicabilidad y la responsabilidad en materia de contratación pública.