The Instrument has to:
— be based on a dry superconducting magnet system i.e. a system that is cooled without the use of liquid helium;
— be able to provide adjustable sample fields of up to 5 Tesla (or more) and sample temperatures (within+/- 1 Kelvin) from room temperature Down to 5 Kelvin;
— be essentially free of vibrations to the sample (the line-broadening from the magnetic system, including the cooling, has to be less than 0,02 mm/second);
— have a design that matches the design of existing Mössbauer spectroscopy equipment at DTU.
38000000 - Laboratorieudstyr, optisk udstyr og præcisionsudstyr (ikke briller)
Postadresse: 11124 South Willow Street
By: Manchester
Postnummer: NH 031303
Land: USA
Det indhold, der offentliggøres på denne side, er udelukkende tænkt som en supplerende tjeneste og har ingen retsvirkning. EU's institutioner påtager sig intet ansvar for dens indhold. De officielle udgaver af de relevante udbudsbekendtgørelser er dem, der er offentliggjort i tillægget til Den Europæiske Unions Tidende og findes i TED. Disse officielle tekster er tilgængelige direkte via linkene på denne side. Yderligere oplysninger findes i meddelelsen om udbudsbetingelser og ansvar.